Title of article :
Accuracy and precision of angiographic volumetry methods for left and right ventricle
Author/Authors :
J. Beier، نويسنده , , E. Wellnhofer، نويسنده , , H. Oswald، نويسنده , , David E. Fleck، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
10
From page :
179
To page :
188
Abstract :
We imaged and quantified 60 ventricle casts (30 LV, 30 RV) to evaluate the accuracy and reliability of angiographic ventricle volumetry. We analyzed the seven biplane methods most frequently used in clinical routine: Arcilla, Arvidsson, Dodge, Ferlinz, Simpson (LV + RV) and Wynne. The ventricle contours were defined by (1) manual drawing on the computer screen, (2) manual drawing using a graphical tablet and (3) automatic contour detection. A high inter-class variation in volume accuracy between the different methods was observed (S.D. = 12.7 ml). The volume methods for the LV (mean differences MDLV: [−2.2, +8.5] ml, average MDLV = 1.8 ml) are more accurate than for the RV (MDRV: [−11.4, +33.1] ml, average MDRV = 12.1 ml). The intrinsic error is about the same for all approaches and is very high: average S.D. = 20 ml, RMS = 185 ml. Manual contour definition results in a volume over-estimation (average MDman = +32.8 ml, r = 0.731) compared with automatic contour detection (average MDauto = +6.2 ml, r = 0.810). LV hypertrophy results in a volume under-estimation of the LV (MDLV = −7 ml) and an over-estimation of the RV (MDRV = +6 ml). RV hypertrophy leads to the opposite effect. It was shown that ventricle volumetry and the calculation of derived parameters (ejection fraction) is extremely case dependent and can only be an estimate of the actual value.
Keywords :
Casts studies , Ventricle volumetry , Biplane angiography , Clinical cardiology
Journal title :
International Journal of Cardiology
Serial Year :
1995
Journal title :
International Journal of Cardiology
Record number :
812051
Link To Document :
بازگشت