• Title of article

    Lost and found: An unusual late complication of the Angio-Seal closure device

  • Author/Authors

    Felicita Andreotti، نويسنده , , Alberto Lavorgna، نويسنده , , Giulio Coluzzi، نويسنده , , Claudio Vincenzoni، نويسنده , , Antonio G. Rebuzzi، نويسنده , , Italo Porto، نويسنده , , Filippo Crea، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    3
  • From page
    1
  • To page
    3
  • Abstract
    We report an unusual case of breakage and embolisation of a femoral access closure device (Angio-Seal), which caused acute leg ischemia after percutaneous coronary intervention in a diabetic patient with documented peripheral vascular disease. Embolectomy from the superficial femoral artery revealed the polymer anchor split in two parts and embedded within a large (25 × 5 mm) thrombus.
  • Keywords
    Breakage of vascular closure device , Acute leg ischemia , peripheral artery disease