• Title of article

    Inappropriate shock delivery by implantable cardioverter defibrillator due to electrical interference with washing machine

  • Author/Authors

    Dhanaraj Singh Chongtham، نويسنده , , Ajay Bahl، نويسنده , , Rohit Manoj Kumar، نويسنده , , K.K. Talwar، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    2
  • From page
    44
  • To page
    45
  • Abstract
    We report a patient with hypertrophic cardiomyopathy who received an inappropriate implantable cardioverter defibrillator shock due to electrical interference from a washing machine. This electrical interference was detected as an episode of ventricular fibrillation with delivery of shock without warning symptoms.
  • Keywords
    Implantable cardioverter defibrillator , washing machine , electrical interference , inappropriate shock
  • Journal title
    International Journal of Cardiology
  • Serial Year
    2007
  • Journal title
    International Journal of Cardiology
  • Record number