Title of article :
Assessment of the distal anastomosis of coronary artery bypass grafts with a 2D T2-weighted turbo spin echo sequence and correlation to conventional coronary angiography
Author/Authors :
Th. Wittlinger، نويسنده , , I. Martinovic، نويسنده , , A. Moritz، نويسنده , , P. Kalden، نويسنده , , K.F. Kreitner، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
Objective
The aim of the study was to evaluate the patensies of the distal anastomoses of coronary artery bypass grafts and to detect graft stenoses and occlusions with a magnetic resonance (MR) spin echo sequence.
Patients and methods
One hundred and eighty-five patients with 481 distal anastomoses were examined with a 1.5 T MR scanner and coronary angiography. A 2-dimensional T2-weigthed breath-hold half-Fourier acquisition single-shot turbo spin echo sequence (Haste) was performed. All images were evaluated independently by a radiologist and cardiologist and compared to the conventional coronary angiography. The observers were blinded to the coronary angiography findings, but informed in regard to the surgical graft anastomosis.
Results
With the Haste sequence, 76% of the distal anastomoses were recognized (368/481). Forty-five of 52 (87%) stenoses and all occlusions were identified. The sensitivity and specificity for the evaluation of the distal anastomosis with the spin echo sequence was 87% and 96%. Twenty-four percent of the distal anastomoses were not identified due to a poor image quality or motion artefacts.
Conclusion
Using the Haste sequence, a reliable assessment of graft patency of the distal anastomosis is possible. Further improvements of the spatial resolution and the image quality are necessary to recommend this MR technique for routine clinical use.
Keywords :
Haste , Coronary artery bypass grafts , magnetic resonance imaging , sequence
Journal title :
International Journal of Cardiology
Journal title :
International Journal of Cardiology