Title of article :
Accidental sizing of patent foramen ovale during atrial septal defect sizing for device
Author/Authors :
N. Krishnakumar، نويسنده , , S. Sivasankaran، نويسنده , , S. Harikrishnan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
2
From page :
40
To page :
41
Abstract :
A 21 year old lady with a 22 mm secundum atrial septal defect with adequate rims, and thin septum at the region of the fossa ovalis, but no additional atrial septal defect by transthoracic and transoesophageal echocardiogram was taken up for ASD sizing with a view to device closure. Initially, the ASD was sized to only 8 mm. However, the distal balloon was not fully opened out, and seemed to be conical with restricted opening. Inadvertent passage of sizing balloon through a patent formen ovale(PFO) was immediately confirmed by transthoracic echocardiogram. The ASD was then recrossed and sized to 24 mm with normal opening out of the distal balloon. This was confirmed by transthoracic echocardiogram. A 28 mm ASD device was deployed across the ASD under transthoracic echocardiographic guidance with no residual flow.
Keywords :
Patent foramen ovale , Atrial septal defect sizing , Atrial septal defect device
Journal title :
International Journal of Cardiology
Serial Year :
2006
Journal title :
International Journal of Cardiology
Record number :
827207
Link To Document :
بازگشت