Title of article :
The magnitude of errors associated in measuring the loads applied on an assistive device While Walking
Author/Authors :
Karimi، Mohammad Taghi نويسنده , , Jamshidi ، Nima نويسنده Department of Biomedical Engineering ,
Issue Information :
فصلنامه با شماره پیاپی 0 سال 2012
Abstract :
Measurement of the loads exerted on the limb is a fundamental part of designing of an assistive device, which has been done by
using strain gauges or a transducer. Although calculation of loads applied on an orthosis coefficients achieved from calibration is a
standard way, most of researchers determined the loads based on available equations. Therefore, the aim of this research is finding the
accuracy of this method with respect to calibration. Some strain gauges were attached on the lateral bar of a reciprocal gait orthosis.
It was calibrated for axial force and the anteroposterior and mediolateral moments. The outputs of strain gauge were changed to
force, and moments based on the coefficients were achieved from calibration using available equations, while 5 normal subjects were
walking with the orthosis. There was a significant difference between the force and moments exerted on the orthosis during walking
based on the two methods (mostly extension moment and compression force). The results of this research indicated that the output of
strain gauge varies based on the methods of calculation. As calibration the gauge is a standard method, it is recommended to calibrate
it before use.
Journal title :
Journal of Medical Signals and Sensors (JMSS)
Journal title :
Journal of Medical Signals and Sensors (JMSS)