Title of article :
Brain lesion detectability studies with a high resolution PET operating in no-septa and partial-septa configurations
Author/Authors :
WANG، Yu نويسنده , , J.، Uribe, نويسنده , , Li، Hongdi نويسنده , , Xing، Tao نويسنده , , Liu، Yaqiang نويسنده , , H.، Baghaei, نويسنده , , M.، Aykac, نويسنده , , Wong، Wai-Hoi نويسنده , , R.، Farrell, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-1363
From page :
1364
To page :
0
Abstract :
Three-dimensional (3-D) positron emission tomography (PET) acquisition in comparison to two-dimensional (2-D) PET acquisition improves the sensitivity of the system at the cost of higher scatter and accidental coincidence contributions. A partial-septa system (i.e., some detector rings lack septa) allowing "partial" 3-D acquisition may provide a better alternative to the 2-D acquisition and full 3-D acquisition. We investigated the effect of partial-septa configuration on the noise equivalent sensitivity and lesion detectability for a high-resolution PET camera. We used the MDAPET camera to detect small lesions in brain images obtained from scan of the Hoffman brain phantom. For this work, three small lesion phantoms with diameters of 3, 5, and 8.6 mm were embedded into the 3-D Hoffman brain phantom. The activity concentration ratio of the lesions to the brain gray matter ranged from 1.5 to 10.
Keywords :
Determination , Continuous , Biodegradable dissolved organic carbon , Cell immobilization , bioreactor
Journal title :
IEEE Transactions on Nuclear Science
Serial Year :
2003
Journal title :
IEEE Transactions on Nuclear Science
Record number :
86208
Link To Document :
بازگشت