Title of article :
Wired collimator for single photon emitter
Author/Authors :
J.، Kato, نويسنده , , K.، Ogawa, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
The aim of this study is to develop a new light collimator with almost the same efficiency in terms of photon penetration as that of the conventional collimator. Generally, the conventional collimator is an over specification in terms of the path length in the material forming the collimator. If the path length of a gamma-ray in the septum is short, many photons may pass through the septum and the septum does not work as a shielding material. On the other hand, if we consider a path located near the center of the septum, it is sufficient to shield from undesirable gamma-rays. So we gave up using the plane-septum structure that is used in the conventional collimator. In the new collimator we made several spaces where the septa located in the conventional collimator. One realistic form of the collimator is composed of many rods whose cross-section is square or circular. The collimator is composed of rod-shaped lead or tungsten in x- and y-directions alternately with spacing in the z-direction (direction of thickness). Even though we align many rods alternately without spacing in z-direction, we can achieve a collimator with less than half the weight of a conventional collimator. Performance of the proposed collimator was confirmed by Monte Carlo simulation and experiments with a handmade collimator.
Keywords :
Biodegradable dissolved organic carbon , Cell immobilization , bioreactor , Determination , Continuous
Journal title :
IEEE Transactions on Nuclear Science
Journal title :
IEEE Transactions on Nuclear Science