Title of article :
Analysis of position-dependent Compton scatter in scintimammography with mild compression
Author/Authors :
S.، Majewski, نويسنده , , M.B.، Williams, نويسنده , , M.J.، More, نويسنده , , D.، Narayanan, نويسنده , , D.A.، Kieper, نويسنده , , P.J.، Goodale, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-1642
From page :
1643
To page :
0
Abstract :
In breast scintigraphy using /sup 99m/Tc-sestamibi the relatively low radiotracer uptake in the breast compared to that in other organs such as the heart results in a large fraction of the detected events being Compton scattered gamma-rays. In this study, our goal was to determine whether generalized conclusions regarding scatter-to-primary ratios at various locations within the breast image are possible, and if so, to use them to make explicit scatter corrections to the breast scintigrams. Energy spectra were obtained from patient scans for contiguous regions of interest (ROIs) centered left to right within the image of the breast, and extending from the chest wall edge of the image to the anterior edge. An anthropomorphic torso phantom with fillable internal organs and a compressed-shape breast containing water only was used to obtain realistic position-dependent scatter-only spectra. For each ROI, the measured patient energy spectrum was fitted with a linear combination of the scatter-only spectrum from the anthropomorphic phantom and the scatter-free spectrum from a point source. We found that although there is a very strong dependence on location within the breast of the scatter-to-primary ratio, the spectra are well modeled by a linear combination of position-dependent scatter-only spectra and a position-independent scatter-free spectrum, resulting in a set of position-dependent correction factors. These correction factors can be used along with measured emission spectra from a given breast to correct for the Compton scatter in the scintigrams. However, the large variation among patients in the magnitude of the position-dependent scatter makes the success of universal correction approaches unlikely.
Keywords :
Cell immobilization , Continuous , bioreactor , Determination , Biodegradable dissolved organic carbon
Journal title :
IEEE Transactions on Nuclear Science
Serial Year :
2003
Journal title :
IEEE Transactions on Nuclear Science
Record number :
86253
Link To Document :
بازگشت