• Title of article

    Detective quantum efficiency of the Medipix pixel detector

  • Author/Authors

    D.W.، Davidson, نويسنده , , J.، Watt, نويسنده , , L.، Tlustos, نويسنده , , B.، Mikulec, نويسنده , , M.، Campbell, نويسنده , , K.، Mathieson, نويسنده , , V.، OShea, نويسنده , , K.M.، Smith, نويسنده , , M.، Rahman, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -1658
  • From page
    1659
  • To page
    0
  • Abstract
    We have measured the intrinsic performance of a digital X-ray detector, the Medipix1, by examining the total detective quantum efficiency (DQE). We studied how the DQE depends on both the incident photon energy and spatial frequency. Reported here is the calculation of the detective quantum efficiency for the case of a 300 (mu)m thick silicon diode detector attached to the Medipix1 readout chip. This was done by determining the modulation transfer function and the noise power spectrum; together these allow the frequency component of the DQE to be calculated. The X-ray absorption efficiency in the detector gives the dependence on incident energy. This system was found to have a DQE that peaked at 0.118, using a dental X-ray source, and dropped to 0.049 at the Nyquist frequency of 2.94 line pairs per mm.
  • Keywords
    Determination , Continuous , Biodegradable dissolved organic carbon , Cell immobilization , bioreactor
  • Journal title
    IEEE Transactions on Nuclear Science
  • Serial Year
    2003
  • Journal title
    IEEE Transactions on Nuclear Science
  • Record number

    86256