Title of article
LET dependence of the charge collection efficiency of silicon microdosimeters
Author/Authors
A.B.، Rosenfeld, نويسنده , , I.M.، Cornelius, نويسنده , , R.، Siegele, نويسنده , , D.D.، Cohen, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-2372
From page
2373
To page
0
Abstract
A heavy ion microprobe was used to conduct ion beam induced charge (IBIC) collection imaging of silicon microdosimeters. The GEANT4 Monte Carlo toolkit was used to simulate these measurements to calculate ion energy loss in the device overlayer and energy deposition in the device sensitive volume. A comparison between experimental and theoretical results facilitated the calculation of charge collection efficiency profiles for several ions.
Keywords
Determination , Continuous , Biodegradable dissolved organic carbon , Cell immobilization , bioreactor
Journal title
IEEE Transactions on Nuclear Science
Serial Year
2003
Journal title
IEEE Transactions on Nuclear Science
Record number
86362
Link To Document