Title of article :
Development of a cesium atomic fountain frequency standard
Author/Authors :
Park، Sang-Eon نويسنده , , Lee، Ho Seong نويسنده , , Kwon، Taeg Yong نويسنده , , Yang، Sung Hoon نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-262
From page :
263
To page :
0
Abstract :
We report on the development of a cesium atomic fountain frequency standard at the Korea Research Institute of Standards and Science (KRISS). Cesium atoms were launched upwards by using an optical moving molasses method after they were cooled in a magnetooptical trap. The temperature of launched atoms was about 2.5 (mu)K. From our cesium atomic fountain, we were able to observe Ramsey fringes of 1-Hz linewidth on the 9.2-GHz clock transition of cesium atoms.