• Title of article

    Systematic shift uncertainty evaluation of IEN CSF1 primary frequency standard

  • Author/Authors

    F.، Levi, نويسنده , , L.، Lorini, نويسنده , , D.، Calonico, نويسنده , , A.، Godone, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -266
  • From page
    267
  • To page
    0
  • Abstract
    We report a preliminary systematic shift uncertainty evaluation of our cesium fountain and of the first comparison with the NIST-F1 clock of the National Institute of Standard and Technology (NIST) with a two way satellite comparison system. The stability of the time scale UTC (IEN), (generated with a commercial Cs beam) is insufficient to allow evaluation of systematic shifts by conventional techniques; therefore, the accuracy evaluation of the fountain was accomplished using mainly fast differential techniques, resulting in an overall uncertainty of 3* 10^-15.