Systematic shift uncertainty evaluation of IEN CSF1 primary frequency standard
Author/Authors
F.، Levi, نويسنده , , L.، Lorini, نويسنده , , D.، Calonico, نويسنده , , A.، Godone, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-266
From page
267
To page
0
Abstract
We report a preliminary systematic shift uncertainty evaluation of our cesium fountain and of the first comparison with the NIST-F1 clock of the National Institute of Standard and Technology (NIST) with a two way satellite comparison system. The stability of the time scale UTC (IEN), (generated with a commercial Cs beam) is insufficient to allow evaluation of systematic shifts by conventional techniques; therefore, the accuracy evaluation of the fountain was accomplished using mainly fast differential techniques, resulting in an overall uncertainty of 3* 10^-15.