Title of article
Measurement of thermoelectric effects in a thermal converter using a NbN/TiN/NbN Josephson junction array
Author/Authors
H.، Sasaki, نويسنده , , K.، Takahashi, نويسنده , , H.، Yamamori, نويسنده , , H.، Fujiki, نويسنده , , A.، Shoji, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-358
From page
359
To page
0
Abstract
The AC-DC transfer difference due to thermoelectric effects in a thermal converter was evaluated using a NbN/TiN/NbN Josephson junction array as an ultra-stable fast-reversed DC (FRDC) source. The Josephson junction array was connected in parallel with a conventional semiconductor FRDC current source, as an "add-on" voltage-stabilizer for the input of the thermal converter. The AC-DC difference due to thermoelectric effects was evaluated with a standard measurement (type-A) uncertainty of the order of 0.01 (mu)V/V, four-times better than that with the conventional semiconductor FRDC source.