Title of article :
A traceable precision current source for currents between 100 aA and 10 pA
Author/Authors :
G.-D.، Willenberg, نويسنده , , H.N.، Tauscher, نويسنده , , P.، Warnecke, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-435
From page :
436
To page :
0
Abstract :
A new calibration setup for picoamperemeters is presented. Owing to the use of a novel voltage ramp generator with increased linearity, the useful measuring time is increased to about 35 min, which reduces the calibration uncertainty of picomaperemeters with long time constants.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91501
Link To Document :
بازگشت