    DC voltage divider calibration system at NMIJ

  • Author/Authors

    H.، Fujiki, نويسنده , , Y.، Sakamoto, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -464
  • From page
    465
  • To page
    0
  • Abstract
    A system to self-calibrate a DC voltage divider with increment taps up to 1 kV has been developed. During selfcalibration, the divider is grounded and is powered at the same level as it is when in normal use. A reference divider self-calibrated by this method is used as a standard to calibrate other dividers at NMIJ. Typical relative uncertainty is 3*10^-7 for 1000 V/10 V and 2*10^-7 for 100 V/10 V (k=2, confidence level of 95%).