Title of article
Analysis of different measurement setups for a programmable Josephson voltage standard
Author/Authors
J.-P.، Lo-Hive, نويسنده , , S.، Djordjevic, نويسنده , , R.، Behr, نويسنده , , J.، Kohlmann, نويسنده , , J.-T.B.M.، Janssen, نويسنده , , P.، Kleinschmidt, نويسنده , , J.M.، Williams, نويسنده , , F.، Piquemal, نويسنده , , P.-O.، Hetland, نويسنده , , D.، Reymann, نويسنده , , G.، Eklund, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-523
From page
524
To page
0
Abstract
The electrical characteristics of two different 1-V binary programmable Josephson arrays, an superconductor/insulator/normal conductor/insulator/superconductor-type Josephson array, and an externally shunted superconductor/insulator/superconductor-type Josephson array, were investigated at ten metrology institutes. Various operational parameters were evaluated and compared using different Josephson array voltage standard setups at microwave frequencies around 70 GHz. The results of the measurements show that both arrays have been working very well and the main differences were not imposed by the arrays themselves, but by the different measurement setups of the laboratories.
Keywords
leukemia
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year
2003
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number
91522
Link To Document