Title of article :
Measurement of C and tan (delta) of a capacitor employing PSDs and dual-slope DVMs
Author/Authors :
V.J.، Kumar, نويسنده , , P.، Sankaran, نويسنده , , K.S.، Rao, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
-1587
From page :
1588
To page :
0
Abstract :
A scheme for measuring the C and tan(delta) (dissipation factor D) values of a capacitor is described in this paper. Comparing the unknown capacitor to a standard capacitor and employing three-phase sensitive detectors and two digital voltmeters, the parameters of the unknown capacitor are determined. The method facilitates testing of a given capacitor at a desired frequency as well as voltage. The feasibility of the scheme is established by building a prototype. The accuracy of the prototype unit for the measurement of C is +-0.2% and that for tan(delta) is +-2.5%.
Keywords :
leukemia
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year :
2003
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number :
91682
Link To Document :
بازگشت