Title of article :
Sources of uncertainty in the nose-to-nose sampler calibration method
Author/Authors :
N.G.، Paulter, نويسنده , , D.R.، Larson, نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
We analyze the nose-to-nose (ntn) method for use as an accurate sampler calibration method. The variations in the measurement of the sampler impulse response using the ntn method are presented, and the validity of the assertion that the kick-out pulse is identical to the sampler impulse response is assessed. Temperature effects on the ntn method are also examined. Finally, the impact of uncertainties in the ntn calibration method on the uncertainties in reported pulse parameters is examined.
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Journal title :
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT