Abstract :
Managing calibration intervals in order to fulfill requirements of quality-oriented organizations is discussed in this paper. In particular, the simple response method is considered. According to this technique, the interval between successive calibrations is adjusted adaptively on the basis of the outcome of the last calibration process. An original analysis is presented about the properties of such calibration interval handling methods. Finally, directions are given on how to tune its parameters optimally, in order to meet desired levels of long-term measurement reliability.