Title of article
Performance of high-T/sub C/ SQUID magnetometers for application in biomagnetism
Author/Authors
Kim، In-Seon نويسنده , , K.K.، Yu, نويسنده , , Y.H.، Lee, نويسنده , , H.C.، Kwon, نويسنده , , Y.K.، Park, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-351
From page
352
To page
0
Abstract
Single layer direct-coupled YBCO SQUID magnetometers have been fabricated and characterized for the purpose of MCG measurements in magnetically disturbed environment. Two types of magnetometers have been designed and fabricated using 10 mm*10 mm substrates. We could operate the conventional 3-mm-wide solid pickup loop magnetometers more stably than the 12-parallel-line pickup loop magnetometers in laboratory environment. We developed a first-order electronic gradiometer system with the SQUID magnetometers with axial displacement of 80 mm without any mechanical alignment of magnetometers. The system with a software filter using calculation of discrete Fourier transform could record clear magnetocardiogram in a weak magnetic-shield-room.
Keywords
homocysteine , Cretan Mediterranean diet , folate , Ischaemic heart disease
Journal title
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY
Serial Year
2003
Journal title
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY
Record number
94040
Link To Document