Title of article :
monte carlo study of fetal dosimetry parameters for 6 mv photon beam
Author/Authors :
atarod، maryam نويسنده Department of Medical Physics, School of Medicine , , shokrani، parvaneh نويسنده Department of Medical Physics, School of Medicine ,
Issue Information :
فصلنامه با شماره پیاپی سال 2013
Abstract :
Because of the adverse effects of ionizing radiation on fetuses, prior to radiotherapy of pregnant patients, fetal dose should be
estimated. Fetal dose has been studied by several authors in different depths in phantoms with various abdomen thicknesses (ATs). In
this study, the effect of maternal AT and depth in fetal dosimetry was investigated, using peripheral dose (PD) distribution evaluations.
A BEAMnrc model of Oncor linac using out of beam components was used for dose calculations in out of field border. A 6 MV photon
beam was used to irradiate a chest phantom. Measurements were done using EBT2 radiochromic film in a RW3 phantom as abdomen.
The followings were measured for different ATs: Depth PD profiles at two distances from the field’s edge, and inplane PD profiles at
two depths. The results of this study show that PD is depth dependent near the field’s edge. The increase in AT does not change PD
depth of maximum and its distribution as a function of distance from the field’s edge. It is concluded that estimating the maximum fetal
dose, using a flat phantom, i.e., without taking into account the AT, is possible. Furthermore, an inplane profile measured at any depth
can represent the dose variation as a function of distance. However, in order to estimate the maximum PD the depth of D
in out of
field should be used for inplane profile measurement.
Journal title :
Journal of Medical Signals and Sensors (JMSS)
Journal title :
Journal of Medical Signals and Sensors (JMSS)