Title of article :
Evaluation of a newly developed diffusion denuder for atmospheric aerosol separation from co-pollutant gases Original Research Article
Author/Authors :
Vaishali Ashok، نويسنده , , Tarun Gupta، نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2012
Pages :
8
From page :
150
To page :
157
Abstract :
This paper presents the design, development and performance evaluation of a non-selective membrane based single annular tube diffusion denuder employed to separate co-pollutant gases from the aerosol of interest. The experiments were carried out on a single annular tube diffusion denuder and the gas removal efficiencies were experimentally determined using a laboratory scale setup as well as under ambient conditions. Using a simple combustion source, the removal efficiency of carbon monoxide was calculated for different main flow (MF) and flush flow (FF) combinations and was found to be in the range of 94–100%. The overall soot particle transmission efficiency was found to be in the range of 70–99% for particle size between 0.3 μm and 2 μm. Overall, transmission efficiencies for submicron particles were also evaluated using an ultrafine condensation particle counter. After the controlled laboratory testing of the prototype denuder, experiments were conducted using ambient air as the test aerosol. This prototype denuder provided a much higher efficiency of gaseous removal in a single stage as compared to earlier reported values. Its simple and easy operation makes it a promising device to use as compared with other commonly employed chemical adsorption based alternate tools.
Keywords :
Non-selective membrane , Sampling artifact , Annular diffusion denuder , Carbon monoxide , Combustion aerosol
Journal title :
Science of the Total Environment
Serial Year :
2012
Journal title :
Science of the Total Environment
Record number :
988499
Link To Document :
بازگشت