Title of article :
Mid-infrared matrix assisted laser desorption ionization with a
waterrglycerol matrix
Author/Authors :
Kathleen L. Caldwell، نويسنده , , Kermit K. Murray )، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
Matrix-assisted laser desorption ionization MALDI.mass spectra were obtained using a water and glycerol matrix with a
tunable mid-infrared optical parametric oscillator. The matrix consists of a 1:1 mixture of water and glycerol deposited on a
thin layer of nitrocellulose and cooled to y308C. When exposed to vacuum, most of the water evaporates, leaving a matrix
of glycerol with residual water. The peptide bradykinin and the protein bovine insulin were used to test this new matrix.
Mass spectra were obtained for bradykinin between 2.76 and 3.1 mm with the maximum analyte signal at 2.8 mm. Mass
resolution in excess of 2000 for bradykinin and 500 for insulin was obtained with delayed ion extraction and a linear time of
flight mass spectrometer. The addition of nitrocellulose to the matrix resulted in exceptionally durable samples: more than
10,000 laser shots which produced analyte signal could be obtained from a single sample spot. q1998 Elsevier Science B.V.
Keywords :
MALDI , matrix , Infrared , mass spectrometry , Laser Desorption
Journal title :
Applied Surface Science
Journal title :
Applied Surface Science