Title of article :
How to measure energy dissipation in dynamic mode atomic
force microscopy
Author/Authors :
B. Anczykowski، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
When studying a mechanical system like an atomic force microscope AFM.in dynamic mode it is intuitive and
instructive to analyse the forces involved in tip–sample interaction. A different but complementary approach is based on
analysing the energy that is dissipated when the tip periodically interacts with the sample surface. This method does not
require solving the differential equation of motion for the oscillating cantilever, but is based entirely on the analysis of the
energy flow in and out of the dynamic system. Therefore the problem of finding a realistic model to describe the tip–sample
interaction in terms of non-linear force–distance dependencies and damping effects is omitted. Instead, it is possible to
determine the energy dissipated by the tip–sample interaction directly by measuring such quantities as oscillation amplitude,
frequency, phase shift and drive amplitude. The method proved to be important when interpreting phase data obtained in
tapping mode, but is also applicable to a variety of scanning probe microscopes operating in different dynamic modes.
Additional electronics were designed to allow a direct mapping of local energy dissipation while scanning a sample surface.
By applying this technique to the cross-section of a polymer blend a material specific contrast was observed. q1999
Elsevier Science B.V. All rights reserved.
Keywords :
Tapping mode , Dynamic mode , Tip–sample interaction , polymer blends , Atomic Force Microscope , energy dissipation
Journal title :
Applied Surface Science
Journal title :
Applied Surface Science