• Title of article

    Effect of step function-like perturbation on intermittent contact mode sensors: a response analysis

  • Author/Authors

    J?nos Kokavecz، نويسنده , , Péter Heszler، نويسنده , , Zsolt T?th، نويسنده , , ?d?m Mechler، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    5
  • From page
    123
  • To page
    127
  • Abstract
    The dynamics of the intermittent contact mode (ICM) probe was investigated. In the experimental study we applied a step function signal to the Z piezo drive and recorded the amplitude signal of the probe while the probe was engaged with the surface. Transient overshoots appear at the edges of the steps. These transients are absent from the control contact force measurements, that is, they are proper to the ICM operation. The phenomenon was investigated by numerical calculations, focused on the effect of change of the drive frequency and the quality factor. We concluded, that the low value of the quality factor results in small transients and short settling time, which are necessary for fast atomic force microscopic operation. Simultaneously, the interaction force increases. Our calculations indicate that the tip–sample force can be lowered by setting the drive frequency slightly below the resonance.
  • Keywords
    Numerical simulation , Fast scanning atomic force microscopy , Intermittent contact mode
  • Journal title
    Applied Surface Science
  • Serial Year
    2003
  • Journal title
    Applied Surface Science
  • Record number

    999904