The development of piezoelectric (PZ) quartz crystal and surface acoustic wave (SAW) biosensors based on fullerene C60 and
immobilized C60-enzymes/antibodies/proteins for the detection of various biological species are reported. The C60 coated
piezoelectric crystal sensors can be applied to the study of interactions between fullerene C60 and some biological species, such
as enzymes, antibodies, proteins and heparin. The partial irreversible responses for some biospecies from C60 molecules were
observed by the desorption study which implied that C60 could chemically react with these biological species. Thus, immobilized
biological species, e.g. C60-GOD, C60-catalase, C60-urease, C60-lipase, C60-anti IgG, C60-heparin, C60-Hb, C60-Mb and
C60-anti-Hb were successfully prepared. The immobilized C60-GOD, C60-catalase, C60-urease, C60-anti-IgG and C60-anti-Hb
were employed as adsorbents onto quartz crystal of various piezoelectric biosensors to detect glucose, H2O2, urea, IgG, and
hemoglobin respectively. The immobilized C60-lipase was applied to distinguishably catalyze the hydrolysis of some optical
isomers such as L- and D-phenyalanine methyl ester and to determine these optical isomers. The immobilized C60-heparin was
employed as a good inhibitor for blood clotting like solvated heparin. The H2O2 bio-sensor was set up with the immobilized
C60-catalase to detect oxygen, the product of the hydrolysis of H2O2 by C60-catalase. The immobilized C60-GOD enzyme
piezoelectric glucose sensor exhibited a good sensitivity and a good lower limit for glucose. A piezoelectric crystal urea biosensor
based on immobilized C60-urease was also prepared to detect urea. Comparison between solvated and immobilized enzymes used
for biosensors was also made. The C60-anti IgG or C60-anti-Hb coated IgG piezoelectric crystal sensors exhibited good
sensitivity, selectivity and repeatability for IgG or hemoglobin. Fullerene C60-Hb and C60-myoglobin (C60-Mb) coated surface
acoustic wave (SAW) immunosensors were prepared to detect the anti-hemoglobin (anti-Hb) and anti-myoglobin (anti-Mb)
antibody, respectively. An electrochemical SAW (ESAW) detection system was also developed to detect glucose in aqueous
solutions.