شماره ركورد كنفرانس :
3961
عنوان مقاله :
Determination of Lead Ion by a Modified Carbon Paste Electrode Based on SiO₂ Nanoparticles, Ionic Liquid and Ionophore (N, N , N -Tricyclo Hexil Phosphoric Triamid) by Potentiometric Method
پديدآورندگان :
Vahedi A ashie.hh.vahedi68@gmail.com Department of chemistry, Mashhad Branch, Islamic Azad University , Ebrahimi M ebrachem2007@yahoo.com
Department of chemistry, Mashhad Branch, Islamic Azad University
كليدواژه :
Ion selective electrode , Carbon paste , Ionic liquid , Ionophore , Potentiometer
عنوان كنفرانس :
سومين همايش ملي تكنولوژي هاي نوين در شيمي، پتروشيمي و نانو ايران
چكيده فارسي :
The aim of this study was the determination of lead ion concentration by Carbon Paste Ion Selective electrode. The electrode response characteristics have been investigated.
Material and methods: The Carbon Paste Electrode (CPE) contains Graphite powder, Nanoparticles, Ionophore and Ionic liquid. The potential measurements were carried out with a Potentiometer. The selectivity of the electrode towards different cationic species was evaluated by Separate Solution Method (SSM).
The results: Among the different compositions studied, the carbon paste incorporating 60% graphite powder, 29% Ionic Liquid, 5%Ionophore and 6%Nanoparticles exhibit the best response characteristics. The electrode exhibits a linear response to Pb (II) over a concentration range (1.0×10−8 – 1.0×10−1) with a Nernstian slope of 30.03 MV per decade. It has a short response time (about 5 s). The detection limit of this electrode was 8.0×10−6 M. The potential remains constant in the range of PH=7-11.
Conclusion: The suggested electrode was successfully used as an indicator for potentiometric determination of Pb (II) in water samples. It was characterized by good analytical parameters: low detection limit, high sensitivity, wide measuring range, wide pH range. All selected cations would not affect the selectivity of the present Lead electrode.