شماره ركورد كنفرانس
2734
عنوان مقاله
A Precise Calibration Method for Torsional Resonant Column Apparatus
پديدآورندگان
Jalili J نويسنده , Darvishi M. M نويسنده
تعداد صفحه
8
كليدواژه
Resonant column , encoder , Strain gauge , Piezoelectric
سال انتشار
1395
عنوان كنفرانس
پنجمين همايش بين المللي مهندسي ژئوتكنيك و مكانيك خاك
زبان مدرك
فارسی
چكيده فارسي
Resonant column apparatus is usually calibrated making use of external acceleration or velocity
transducers, according to the related standards. A series experimental study was conducted on the
resonant column apparatus of International Institute of Earthquake Engineering and Seismology, to assure
a precise calibration of the device. In addition to a piezoelectric accelerometer which was used to check
the accuracy of the acceleration transducer of the apparatus, a high-precision encoder was also utilized, to
exactly measure the magnitude and also location of the vibration amplitude measurement. To finally
check the reliability of the precise calibration, local strain gauges were attached to a dummy specimen
and the vibration amplitude measured by the apparatus and also by the strain gauges were measured,
which were satisfactorily similar. The proposed calibration method may be followed in similar cases, as
stated through the paper
شماره مدرك كنفرانس
4327200
سال انتشار
1395
از صفحه
1
تا صفحه
8
سال انتشار
1395
لينک به اين مدرک