• Author/Authors

    İyisoy, Atila Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Kurşaklıoğlu, Hürkan Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Çelik, Turgay Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Barçın, Cem Gulhane Military Medical Faculty - Department of Cardiology, Turkey , Çelik, Murat Gulhane Military Medical Faculty - Department of Cardiology, Turkey

  • Title Of Article

    Retrieval of the Amplatzer duct occluder II device embolizing to the pulmonary artery by Goose Neck snare

  • شماره ركورد
    19865
  • Abstract
    Percutaneous transcatheter closure of patent ductus arteriosus is a wellestablished procedure. Although it has been known as a safe treatment strategy, some complications have been reported. Embolization of the device, especially to the pulmonary artery, is one of these complications. In this report, we present a patient in whom an Amplatzer® duct occluder (ADO) II device embolized to the pulmonary artery and it was retrieved by using “Goose Neck” snare. We would like to notice the great importance of accurate preprocedural measurement of ductus arteriosus diameter during percutaneous transcatheter closure which seems to be a simple procedure.
  • From Page
    276
  • NaturalLanguageKeyword
    Amplatzer duct occluder , “Goose Neck” snare , patent ductus arteriosus , percutaneous transcatheter closure
  • JournalTitle
    Gulhane Medical Journal
  • To Page
    279
  • JournalTitle
    Gulhane Medical Journal