• DocumentCode
    1000777
  • Title

    Comparison of measured responses in two spectrally-sensitive X-ray detectors to predictions obtained using the ITS radiation transport code

  • Author

    Carlson, G.A. ; Beutler, D.E. ; Seager, K.D. ; Knott, D.P.

  • Author_Institution
    Sandia Nat. Labs., Albuquerque, NM, USA
  • Volume
    35
  • Issue
    6
  • fYear
    1988
  • fDate
    12/1/1988 12:00:00 AM
  • Firstpage
    1272
  • Lastpage
    1277
  • Abstract
    Responses of a Ge detector and a filtered thermoluminescent dosimeter (TLD) array detector were measured at a steady-state bremsstrahlung source (the Pelletron), at endpoint energies from 150 to 900 keV. Predictions of detector response using the Integrated Tiger Series (ITS) Monte Carlo codes are found to be in excellent agreement with measured responses for both detectors. The results extend the range of validity of the ITS codes. With calibration provided by these experiments and by ITS predictions, dose-depth data from the TLD arrays can be used to estimate flash X-ray source endpoint energies
  • Keywords
    Monte Carlo methods; X-ray detection and measurement; nuclear engineering computing; semiconductor counters; thermoluminescent dosimeters; 150 to 900 keV; Ge detector; ITS radiation transport code; Integrated Tiger Series; Monte Carlo codes; Pelletron; TLD array detector; calibration; detector response; dose-depth data; endpoint energies; filtered thermoluminescent dosimeter; flash X-ray source; spectrally-sensitive X-ray detectors; steady-state bremsstrahlung source; Collimators; Electron beams; Energy measurement; Germanium; Laboratories; Monte Carlo methods; Radiation detectors; Sensor arrays; Testing; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.25451
  • Filename
    25451