DocumentCode
1000777
Title
Comparison of measured responses in two spectrally-sensitive X-ray detectors to predictions obtained using the ITS radiation transport code
Author
Carlson, G.A. ; Beutler, D.E. ; Seager, K.D. ; Knott, D.P.
Author_Institution
Sandia Nat. Labs., Albuquerque, NM, USA
Volume
35
Issue
6
fYear
1988
fDate
12/1/1988 12:00:00 AM
Firstpage
1272
Lastpage
1277
Abstract
Responses of a Ge detector and a filtered thermoluminescent dosimeter (TLD) array detector were measured at a steady-state bremsstrahlung source (the Pelletron), at endpoint energies from 150 to 900 keV. Predictions of detector response using the Integrated Tiger Series (ITS) Monte Carlo codes are found to be in excellent agreement with measured responses for both detectors. The results extend the range of validity of the ITS codes. With calibration provided by these experiments and by ITS predictions, dose-depth data from the TLD arrays can be used to estimate flash X-ray source endpoint energies
Keywords
Monte Carlo methods; X-ray detection and measurement; nuclear engineering computing; semiconductor counters; thermoluminescent dosimeters; 150 to 900 keV; Ge detector; ITS radiation transport code; Integrated Tiger Series; Monte Carlo codes; Pelletron; TLD array detector; calibration; detector response; dose-depth data; endpoint energies; filtered thermoluminescent dosimeter; flash X-ray source; spectrally-sensitive X-ray detectors; steady-state bremsstrahlung source; Collimators; Electron beams; Energy measurement; Germanium; Laboratories; Monte Carlo methods; Radiation detectors; Sensor arrays; Testing; X-ray detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.25451