DocumentCode
1002794
Title
Parallel acquisition of MR images using time multiplexed coils
Author
Wright, S.W. ; Porter, Jay R.
Author_Institution
Dept. of Electr. Eng., Texas A&M Univ., College Station, TX, USA
Volume
28
Issue
1
fYear
1992
Firstpage
71
Lastpage
72
Abstract
A new technique for parallel acquisition of magnetic resonance images is described. Two RF coils are time multiplexed into a single receiver channel, halving the effective image acquisition time with no degradation in signal-to-noise ratio. Noise filtering is performed with RF bandpass filters. In principle, the method can be extended to additional coils.
Keywords
biomedical NMR; data acquisition; picture processing; RF bandpass filters; RF coils; SNR; magnetic resonance images; noise filtering; parallel acquisition; single receiver channel; time multiplexed; time multiplexed coils;
fLanguage
English
Journal_Title
Electronics Letters
Publisher
iet
ISSN
0013-5194
Type
jour
DOI
10.1049/el:19920044
Filename
255930