• DocumentCode
    1003901
  • Title

    A high frequency iron loss measuring system

  • Author

    Katane, Tamotsu ; Yamaguchi, Takayoshi ; Ito, Tatsushi ; Sato, Toshiro ; Sakaki, Yo

  • Author_Institution
    Chiba University, Chiba-shi, Japan
  • Volume
    20
  • Issue
    5
  • fYear
    1984
  • fDate
    9/1/1984 12:00:00 AM
  • Firstpage
    1729
  • Lastpage
    1731
  • Abstract
    A high performance iron loss measuring system is introduced by using an 8-bit microcomputer, a two-channel memory device with high speed A/D converters and a wide band power amplifier. A required time to obtain a datum of one experimental point is about one minute when the error between the desired maximum flux density and the settled one is allowed within 0.5%. This system makes it possible to measure iron losses in the frequency range up to 1MHz. The results have a good correlation to the measured results by using the U-Function Meter.
  • Keywords
    Iron materials/devices; Magnetic losses; Magnetic measurements; Circuits; Frequency conversion; Frequency measurement; Impedance; Iron; Loss measurement; Microcomputers; Random access memory; Regulators; Voltage;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.1984.1063272
  • Filename