Abstract
Positron emission tomography (PET) requires accurate timing of scintillation events to properly discriminate between coincident and noncoincident pairs. The traditional solution to timing is based on custom application specific integrated circuits (ASIC) designs, whose cost may not be justified in the design of experimental small animal PET scanners. The new generation of PET scanners introduces the idea of continuous sampling of the detected scintillation pulse, replacing event-triggered acquisition front-ends. This approach enables new options to the timing procedure based on digital processing of the sampled pulse signal. This work proposes a time stamping algorithm based on the optically matched filter and compares the potential performance benefits of this approach versus other FIR-based timing algorithms, some of which have been already implemented by different authors. Results show that the coincidence timing resolution may be as low as 1.5 ns without the need of expensive high-speed converters when the proper signal processing is applied.
Keywords
data acquisition; medical signal processing; positron emission tomography; positron emission tomography; pulse signal digital processing; real-time digital timing; scintillation pulse continuous sampling; small animal PET scanners; Data acquisition; positron emission tomography; signal processing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on