DocumentCode
1009427
Title
Digital pulse processing in high resolution, high throughput, gamma-ray spectroscopy
Author
Georgiev, Andrey ; Gast, Werner
Author_Institution
Inst. fuer Kernphys., Forschungszentrum Julich, Germany
Volume
40
Issue
4
fYear
1993
fDate
8/1/1993 12:00:00 AM
Firstpage
770
Lastpage
779
Abstract
A new method for processing signals produced by high resolution, large volume semiconductor detectors is described. These detectors, to be used in the next generation of spectrometer arrays for nuclear research (i.e., EUROBALL, etc.), present a set of problems, such as resolution degradation due to charge trapping and ballistic deficit effects, poor resolution at a high count rate, long term and temperature instability, etc. To solve these problems, a new approach based on digital moving window deconvolution (MWD) is developed
Keywords
digital circuits; gamma-ray spectroscopy; nuclear electronics; signal processing; digital moving window deconvolution; digital pulse processing; gamma-ray spectroscopy; high count rate; high resolution; high throughput; resolution degradation; semiconductor detectors; Deconvolution; Energy resolution; Instruments; Noise shaping; Preamplifiers; Semiconductor device noise; Signal processing; Signal resolution; Spectroscopy; Throughput;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.256659
Filename
256659
Link To Document