DocumentCode :
1010262
Title :
Accurate intensity correction for endorectal surface coil MR imaging of the prostate
Author :
Singh, Manbir ; NessAiver, Moriel
Author_Institution :
Dept. of Radiol. & Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
Volume :
40
Issue :
4
fYear :
1993
fDate :
8/1/1993 12:00:00 AM
Firstpage :
1169
Lastpage :
1173
Abstract :
An accurate technique is developed to correct the intensity variation in the magnetic resonance images acquired with an endorectal surface coil for prostate imaging. The technique relies on reconstructing the in vivo size, shape, and location of the coil during imaging, wherefrom a pixel-by-pixel correction is applied to the coil´s sensitivity profile in accordance with the Biot-Savart law. The method is evaluated by a test-object containing multiple layers of beads embedded in gelatine. Qualitatively, the improvement in visualizing regions of interest after corrections is obvious. Quantitatively, the local SNR shows an improvement by a factor of 1.5-2. Several human studies are performed. The technique shows remarkable improvement in visualizing the prostate region
Keywords :
biomedical NMR; medical image processing; patient diagnosis; Biot-Savart law; endorectal surface coil; in vivo size; magnetic resonance images; pixel-by-pixel correction; prostate imaging; Coils; Image reconstruction; In vivo; Magnetic resonance; Magnetic resonance imaging; Pixel; Shape; Surface reconstruction; Testing; Visualization;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.256731
Filename :
256731
Link To Document :
بازگشت