Title :
The energizing of a NMR superconducting coil by a superconducting rectifier
Author :
Sikkenga, J. ; Knoben, J. ; Spoorenberg, C. J G ; Kraaij, G.J. ; Klundert, L. J M V D ; Kate, H. H J Ten
Author_Institution :
HOLEC Innovation and Technology, Hengelo, The Netherlands
fDate :
3/1/1985 12:00:00 AM
Abstract :
NMR magnets require a good homogeneity within a certain volume and an excellent field stability. The homogeneity can be met using a superconducting shim coil system. The field stability requires a constant current, although in many cases the current decay time constant is too low, due to imperfections in the superconducting wire and joints. This can be overcome using a rectifier. The rectifier can also be used to load the coil. The combination and interaction of the superconducting NMR coil (2.0 Tesla and 0.35 m cold bore) and the rectifier (20 W / 1 kA) is tested. The safety of the system is discussed. The shim coil system can compensate the strayfield of the rectifier. The field decay compensation will be discussed.
Keywords :
Nuclear magnetic resonance imaging; Rectifiers; Superconducting coils; Boring; Nuclear magnetic resonance; Rectifiers; Safety; Stability; Superconducting coils; Superconducting filaments and wires; Superconducting magnets; Switches; Technological innovation;
Journal_Title :
Magnetics, IEEE Transactions on
DOI :
10.1109/TMAG.1985.1063857