DocumentCode
1011228
Title
Dynamic posturography: analysis of error in force plate measurement of postural sway
Author
Barin, Kamran
Author_Institution
Dept. of Otolaryngology, Ohio State Univ., Columbus, OH, USA
Volume
11
Issue
4
fYear
1992
Firstpage
52
Lastpage
56
Abstract
EquiTest, a framework for the analysis of error in dynamic posturography, is described. Results presented indicate that systematic errors may be present when force plate data are used to estimate body sway. The response parameters in EquiTest are derived from measured forces. Two of these parameters, equilibrium score (ES) and strategy score (SS), are analyzed. ES is shown to be affected by the dynamic component of body sway. The analysis of error for SS suggests that this score is a poor indicator of the relative movements of the upper and lower body.<>
Keywords
biomechanics; biomedical measurement; force measurement; measurement errors; EquiTest; dynamic posturography; equilibrium score; force plate measurement; lower body movements; postural sway; strategy score; upper body movements; Control systems; Differential equations; Error analysis; Foot; Force measurement; Hip; Manufacturing; Stability; Testing; Transducers;
fLanguage
English
Journal_Title
Engineering in Medicine and Biology Magazine, IEEE
Publisher
ieee
ISSN
0739-5175
Type
jour
DOI
10.1109/51.256960
Filename
256960