• DocumentCode
    1015775
  • Title

    Stability of a fiber-optic magnetometer

  • Author

    Koo, Kee P. ; Bucholtz, Frank ; Dandridge, A. ; Tveten, A.B.

  • Author_Institution
    Sachs Freeman Associates, Landover, MD
  • Volume
    22
  • Issue
    3
  • fYear
    1986
  • fDate
    5/1/1986 12:00:00 AM
  • Firstpage
    141
  • Lastpage
    144
  • Abstract
    The stability of a fiber-optic magnetometer employing metallic glass ribbons was examined over a period of more than 10 h. The magnetometer can be operated either in the open magnetic loop or the closed magnetic loop configuration. The short-term minimum detectable magnetic field for ≈ 10 cm of sensor fiber length was less than 2 gamma in the dc to 1-Hz bandwidth. Long-term stability of approximately 10 gamma was achieved in the closed magnetic loop configuration.
  • Keywords
    Magnetometers; Magnetostrictive transducers; Optical fiber transducers; Amorphous magnetic materials; Glass; Magnetic anisotropy; Magnetic field induced strain; Magnetic field measurement; Magnetometers; Magnetostriction; Perpendicular magnetic anisotropy; Saturation magnetization; Stability;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/TMAG.1986.1064293
  • Filename
    1064293