DocumentCode
1015775
Title
Stability of a fiber-optic magnetometer
Author
Koo, Kee P. ; Bucholtz, Frank ; Dandridge, A. ; Tveten, A.B.
Author_Institution
Sachs Freeman Associates, Landover, MD
Volume
22
Issue
3
fYear
1986
fDate
5/1/1986 12:00:00 AM
Firstpage
141
Lastpage
144
Abstract
The stability of a fiber-optic magnetometer employing metallic glass ribbons was examined over a period of more than 10 h. The magnetometer can be operated either in the open magnetic loop or the closed magnetic loop configuration. The short-term minimum detectable magnetic field for ≈ 10 cm of sensor fiber length was less than 2 gamma in the dc to 1-Hz bandwidth. Long-term stability of approximately 10 gamma was achieved in the closed magnetic loop configuration.
Keywords
Magnetometers; Magnetostrictive transducers; Optical fiber transducers; Amorphous magnetic materials; Glass; Magnetic anisotropy; Magnetic field induced strain; Magnetic field measurement; Magnetometers; Magnetostriction; Perpendicular magnetic anisotropy; Saturation magnetization; Stability;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/TMAG.1986.1064293
Filename
1064293