• DocumentCode
    1019530
  • Title

    Numerical analysis of eddy current problems involving z gradient coils in superconducting MRI magnets

  • Author

    Takahashi, Takao

  • Author_Institution
    Hitachi Ltd., Ibaraki, Japan
  • Volume
    26
  • Issue
    2
  • fYear
    1990
  • fDate
    3/1/1990 12:00:00 AM
  • Firstpage
    893
  • Lastpage
    896
  • Abstract
    The pulsive field of z gradient coils induces eddy currents in the cryostat walls of MRI (magnetic resonance imaging) superconducting magnets. These eddy currents are computed here by a network method and the effects of the eddy currents are investigated while varying the resistivity of the cryogenic radiation shield. Two paired z gradient coils are proposed to improve the efficiency and linearity of the z field gradient
  • Keywords
    biomedical NMR; coils; eddy currents; superconducting magnets; cryogenic radiation shield; cryostat walls; eddy currents; magnetic resonance imaging; network method; pulsive field; superconducting magnets; z gradient coils; Cryogenics; Eddy currents; Equations; Inductance; Magnetic resonance imaging; Magnetic shielding; Numerical analysis; Steel; Superconducting coils; Superconducting magnets;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.106461
  • Filename
    106461