DocumentCode :
1022922
Title :
Performance Simulation of a Quadrupole Mass Filter Operating in the First and Third Stability Zones
Author :
Hogan, Thomas J. ; Taylor, Stephen
Author_Institution :
Univ. of Liverpool, Liverpool
Volume :
57
Issue :
3
fYear :
2008
fDate :
3/1/2008 12:00:00 AM
Firstpage :
498
Lastpage :
508
Abstract :
A method of computation that accurately simulates the performance of quadrupole mass filters (QMFs) is described. Behavior is described by determining the individual trajectories of a large number of ions (108) as they are injected into the QMF. The effects of the ratio of circular electrode radius r to electric field radius r0 on the performance characteristics have been investigated for zone 1 (a ap 0.237 and q ap 0.706) and zone 3 (a ap 3.16 and q ap 3.23) operation. We demonstrate that performance sensitivity to the r/r0 ratio is different for zone 3 than those previously reported for zone 1. The magnitude and variation of the ldquotailrdquo in the mass spectral peak shapes that are apparent for zone 1 is much decreased for zone 3 and does not influence QMF resolution. Variation in ion trajectories and associated power-spatial frequency spectra when operated in zones 1 and 3 with varying r/r0 geometrical ratios are also presented. We demonstrate that these provide an alternative method in determining an ideal value for r/r0.
Keywords :
filters; geometry; stability; QMF; circular electrode; first stability zones; geometrical ratios; ion trajectories; mass spectral peak shapes; power-spatial frequency spectra; quadrupole mass filter; third stability zones; Mathieu stability regions; multipoles; power frequency spectrum; quadrupole mass filter (QMF); spatial frequencies;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2007.911632
Filename :