DocumentCode :
1026747
Title :
A Tachometer-Accelerometer-Vibrometer for Instrument-Gyroscope Testing
Author :
Laurenson, R.M. ; Long, T.H.
Author_Institution :
Union Switch and Signal Company, Swissvale, Pa.
Volume :
64
Issue :
8
fYear :
1945
Firstpage :
593
Lastpage :
596
Abstract :
This paper gives an account of the development, operating principles, and use of a possibly unique instrument that has been and is of considerable value on instrument-gyroscope manufacturing and repair production lines. It is used to indicate gyroscope-rotor speed when the rotor is brought up to 1,000 rpm for bearing adjustment. Then, by the turn of a switch, the instrument indicates the rate of deceleration as the bearing is adjusted, thus permitting accurate adjustment of bearing pressures. A third position of the same selector switch gives tachometer operation up to 15,000 rpm for virbation tests. Excessive vibration is indicated visibly and audibly through another amplifier, independently of the selector-switch position.
Keywords :
Gyroscopes; Instruments; Pressure measurement; Probes; Rotors; Switches; Synchronous motors; Testing; Time measurement; Velocity measurement;
fLanguage :
English
Journal_Title :
American Institute of Electrical Engineers, Transactions of the
Publisher :
ieee
ISSN :
0096-3860
Type :
jour
DOI :
10.1109/T-AIEE.1945.5059182