DocumentCode
1029425
Title
FastSPECT II: a second-generation high-resolution dynamic SPECT imager
Author
Furenlid, Lars R. ; Wilson, Donald W. ; Chen, Yi-Chun ; Kim, Hyunki ; Pietraski, Philip J. ; Crawford, Michael J. ; Barrett, Harrison H.
Author_Institution
Dept. of Radiol., Univ. of Arizona, Tucson, AZ, USA
Volume
51
Issue
3
fYear
2004
fDate
6/1/2004 12:00:00 AM
Firstpage
631
Lastpage
635
Abstract
FastSPECT II is a recently commissioned 16-camera small-animal SPECT imager built with modular scintillation cameras and list-mode data-acquisition electronics. The instrument is housed in a lead-shielded enclosure and has exchangeable aperture assemblies and adjustable camera positions for selection of magnification, pinhole size, and field of view. The calibration of individual cameras and measurement of an overall system imaging matrix (1 mm3 voxels) are supported via a five-axis motion-control system. Details of the system integration and results of characterization and performance measurements are presented along with first tomographic images. The dynamic imaging capabilities of the instrument are explored and discussed.
Keywords
biomedical imaging; cameras; single photon emission computed tomography; 16-camera small-animal; FastSPECT II; adjustable camera positions; calibration; exchangeable aperture assemblies; field of view selection; five-axis motion-control system; lead-shielded enclosure; list-mode data-acquisition electronics; magnification selection; modular scintillation cameras; overall system imaging matrix; pinhole size selection; second-generation high-resolution dynamic SPECT imager; tomographic images; Animals; Assembly; Biomedical optical imaging; Calibration; Cameras; High-resolution imaging; Instruments; Optical imaging; Photomultipliers; Tomography; Dynamic; SPECT; high-resolution; list-mode; small animal;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2004.830975