DocumentCode
1031268
Title
Proporsal of blood circulation simulator which utilized two linear DC motors
Author
Yamada, H. ; Ota, H. ; Ogasawara, T. ; Watabe, Y.
Author_Institution
Shinshu University, Nagano, Japan
Volume
23
Issue
5
fYear
1987
fDate
9/1/1987 12:00:00 AM
Firstpage
2434
Lastpage
2436
Abstract
A blood circulation simulator is proposed on the assumption that the blood circulation system of living bodies is a system operated by two pumps, the heart pump and the peripheral pump. System configuration is described of a blood circulation simulator for which the two pumps mentioned above are constructed with linear DC motors, and a calculation is made of pulmonary arterial pressure for the operation of both the pumps in synchronous states. A comparison has revealed a similarity of the calculated values of pulmonary arterial pressure to observed values, verifying the applicability of the simulator. The characteristics of the LDMs used for the simulator also were established.
Keywords
Blood flow; DC motors; Linear motors; Arterial blood circulation; Artificial heart; Blood pressure; Blood vessels; DC motors; Pumps; Reservoirs; Testing; Valves; Water resources;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/TMAG.1987.1065670