A widely linear temperature to frequency converter using a thermistor in a pulse generator
Author
Sengupta, R.N.
Author_Institution
Central Sci. Instrum. Org., Chandigarh, India
Volume
37
Issue
1
fYear
1988
fDate
3/1/1988 12:00:00 AM
Firstpage
62
Lastpage
65
Abstract
A modification of an existing technique for thermistor linearization is presented. A temperature-measuring circuit using a thermistor in a specially designed pulse generator is used to obtain a highly linear relation between absolute temperature and frequency. The use of a general-purpose operational amplifier and two switching transistors in the design of the pulse generator provides a temperature scale that is linear over a wide range of values on a high degree of sensitivity. The result shows that an extremely high degree of linearity is achieved by this method
Keywords
convertors; linearisation techniques; pulse generators; temperature measurement; thermistors; general-purpose operational amplifier; linear temperature to frequency converter; linearization; pulse generator; square wave generator; switching transistors; temperature-measuring circuit; thermistor; Bridge circuits; Frequency conversion; Frequency measurement; Linearity; Operational amplifiers; Pulse generation; Temperature distribution; Temperature sensors; Thermistors; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.2665