    Investigation of the energy resolution and charge collection efficiency of Cd(Zn)Te detectors with three electrodes

  • Author

    Kim, Hadong ; Cirignano, Leonard ; Shah, Kanai ; Squillante, Michael ; Wong, Philip

  • Author_Institution
    Radiat. Monitoring Devices Inc., Watertown, MA, USA
  • Volume
    51
  • Issue
    3
  • fYear
    2004
  • fDate
    6/1/2004 12:00:00 AM
  • Firstpage
    1229
  • Lastpage
    1234
  • Abstract
    Radiation Monitoring Devices, Inc., has recently investigated Cd(Zn)Te detector with three electrodes. Planar devices were refabricated as three-electrode devices after recording pulse height spectra. From the spectroscopy measurements at room temperature, photon energy resolutions of 8.3% and 3.5% FWHM were obtained, without noticeable loss of the detector efficiency, with the "three-electrode detector" for gamma ray energies of 122 and 662 keV, respectively, while no discernable full energy peaks were apparent with the planar detectors. The full contact cathode was grounded while two different bias voltages were applied to each anode electrode, even though the induced signal was collected only from the center anode. By investigation of the real and weighting electric field of the detector, the possible reasons for the improved spectra with the three-electrode detector were determined. The strong real electric field around the two anodes very likely enhances the created charge velocity and results in increasing the possibility to induce full charge collection. From additional measurements with increased differences between two bias voltages applied to the anodes, the effects of the strong real electric field on the improved energy spectrum have been confirmed.
  • Keywords
    electric charge; electric fields; electrodes; gamma-ray detection; gamma-ray spectra; nuclear electronics; pulse height analysers; radiation monitoring; semiconductor counters; 293 to 298 K; Cd(Zn)Te detectors; anode; bias voltages; charge collection efficiency; charge velocity; contact cathode; detector efficiency; detector electric field; energy peaks; energy spectrum; gamma ray energies; induced signal; photon energy resolutions; planar detectors; pulse height spectra; radiation monitoring devices; room temperature; spectroscopy measurements; three-electrode devices; Anodes; Electrodes; Energy measurement; Energy resolution; Gamma ray detection; Gamma ray detectors; Radiation detectors; Radiation monitoring; Temperature measurement; Voltage; Author, please supply your own keywords or send a blank e-mail to keywords@ieee.org to receive a list of suggested keywords;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2004.829385
  • Filename
    1312046