Title :
Comb-Generator Characterization
Author :
Reader, Howard Charles ; Williams, Dylan F. ; Hale, Paul D. ; Clement, Tracy S.
Author_Institution :
Stellenbosch Univ., Stellenbosch
Abstract :
We characterize a 50-GHz comb generator with a sampling oscilloscope. With careful control of the input power, input harmonics, and comb generator temperature, we measure the output spectrum with a standard uncertainty of 0.1 dB and 0.5deg. We correct the measurements for time-base distortion, impedance mismatch, an inline attenuator, and the complex frequency response of the oscilloscope´s sampler. We also report on the stability of the comb-generator spectrum and the effects of the recorded time windows, drive levels, and temperature. These results provide general guidelines for practioners of high-speed measurements and a benchmark for future inter-laboratory comparisons of harmonic phase-reference calibrations.
Keywords :
attenuators; frequency response; impedance matching; oscilloscopes; signal generators; signal sampling; comb-generator characterization; complex frequency response; frequency 50 GHz; high-speed measurement; impedance mismatch; inline attenuator; sampling oscilloscope; time-base distortion; Character generation; Distortion measurement; Measurement standards; Oscilloscopes; Power generation; Power measurement; Power system harmonics; Sampling methods; Temperature control; Temperature measurement; Comb-generator stability; mismatch correction; sampling-oscilloscope calibration;
Journal_Title :
Microwave Theory and Techniques, IEEE Transactions on
DOI :
10.1109/TMTT.2007.914630