DocumentCode
1035213
Title
International comparison GT-RF 78-14: power (1 mW) at 50 MHz in 50-Ω coaxial line
Author
Jurkus, Algirdas
Author_Institution
Nat. Res. Council, Ottawa, Ont., Canada
Volume
37
Issue
1
fYear
1988
fDate
3/1/1988 12:00:00 AM
Firstpage
160
Lastpage
162
Abstract
An international comparison of power standards at 50 MHz was organized at the 1978 meeting of the Groupe de Travail pour les Grandeurs aux Radiofrequences. Eleven national laboratories from Canada (pilot laboratory), USA, Japan, Australia, France, FRG, Italy, UK, the Netherlands, Sweden, and Hungary participated. The traveling standards were two commercially available thermistor mounts supplied by the pilot laboratory. The measured values of effective efficiency fall within a range of 0.15% for the majority of participants. Periodic remeasurement at the pilot laboratory showed no evidence of drift over the five-year span of the intercomparison
Keywords
measurement standards; power measurement; 1 mW; 1978; 50 MHz; 50 ohm; Australia; Canada; FRG; France; Hungary; Italy; Japan; Netherlands; Sweden; UK; USA; coaxial line; international comparison; measurement standard; national laboratories; pilot laboratory; power standards; thermistor mounts; traveling standards; Antenna measurements; Australia; Coaxial components; Councils; Frequency; Laboratories; NIST; Particle measurements; Permittivity measurement; Thermistors;