Title :
Design of coils for millimeter- and submillimeter-sized biotelemetry
Author :
Pichorim, Sérgio Francisco ; Abatti, Paulo José
Author_Institution :
CPGEI/CEFET-PR-Parand Fed. Center of Technol. Educ., Parana, Brazil
Abstract :
Biotelemetric systems, especially those that employ implanted devices, work with inductive links, where usually large circular external coils are separated by relatively large distances (dimensions of centimeters) from the small (millimeter- or submillimeter-size) implanted coils. This paper shows that, under these conditions, a simplified method for calculation of the mutual inductance (M) between the coils, avoiding elliptic integrals, can be obtained. A procedure for coil design, with maximum M between them, is also described.
Keywords :
biomedical telemetry; inductance; prosthetics; solenoids; coil design; implantable devices; inductive links; millimeter-sized biotelemetry; mutual inductance; submillimeter-sized biotelemetry; Biomedical telemetry; Coaxial components; Coils; Educational technology; Ferrites; Inductance; Magnetic analysis; Magnetic flux; Magnetic flux density; Solenoids; Computer-Aided Design; Electric Conductivity; Equipment Design; Equipment Failure Analysis; Telemetry;
Journal_Title :
Biomedical Engineering, IEEE Transactions on
DOI :
10.1109/TBME.2004.827542