• DocumentCode
    1045358
  • Title

    A Cylindrical Silicon-on-Insulator Microdosimeter: Charge Collection Characteristics

  • Author

    Ziebell, A.L. ; Lim, W.H. ; Reinhard, M.I. ; Cornelius, I. ; Prokopovich, D.A. ; Siegele, R. ; Dzurak, A.S. ; Rosenfeld, A.B.

  • Author_Institution
    Centre for Med. Radiat. Phys., Univ. of Wollongong, Wollongong, NSW
  • Volume
    55
  • Issue
    6
  • fYear
    2008
  • Firstpage
    3414
  • Lastpage
    3420
  • Abstract
    A novel silicon-on-insulator microdosimeter for estimating the radiobiological effectiveness (RBE) of a mixed radiation field is presented. An ion beam induced charge collection study has confirmed the microdosimeter possesses well defined micron sized 3D cylindrical sensitive volumes. An array of these SVs has the capability of studying the track structure of high energy heavy ions typical of a deep space environment.
  • Keywords
    dosimeters; silicon-on-insulator; charge collection characteristics; cylindrical silicon-on-insulator microdosimeter; ion beam induced charge collection; micron sized 3D cylindrical sensitive volumes; mixed radiation field; radiobiological effectiveness; Australia Council; Instruments; Neutrons; Nuclear and plasma sciences; Radiation detectors; Silicon on insulator technology; Single event upset; Size measurement; Solid state circuits; Volume measurement; Ion beams; microdosimetry; radiobiological effectiveness; silicon-on-insulator;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2008.2004464
  • Filename
    4723726